Experts has a new look! Let us know what you think of the updates.

Provide feedback
Home
Scholarly Works
Prospective Study of Tricuspid Regurgitation...
Conference

Prospective Study of Tricuspid Regurgitation Associated With Permanent Leads After Cardiac Rhythm Device Implantation

Abstract

BACKGROUND: Tricuspid regurgitation (TR) has been associated with cardiac rhythm device (CRD) implantation with intracardiac lead insertion. However, data on the incidence of postdevice TR are limited and largely from retrospective studies. We hypothesized that permanent lead implantation would be associated with an increase in TR. METHODS: We prospectively included consecutive patients with a clinical indication for CRD. Patients underwent …

Authors

Van De Heyning CM; Elbarasi E; Masiero S; Brambatti M; Ghazal S; Al-Maashani S; Capucci A; Leong D; Shivalkar B; Saenen JB

Volume

35

Pagination

pp. 389-395

Publisher

Elsevier

Publication Date

April 2019

DOI

10.1016/j.cjca.2018.11.014

Conference proceedings

Canadian Journal of Cardiology

Issue

4

ISSN

0828-282X