Accuracy of Robotic Radiosurgical Liver Treatment Throughout the Respiratory Cycle

Abstract

PURPOSE: To quantify random uncertainties in robotic radiosurgical treatment of liver lesions with real-time respiratory motion management. METHODS AND MATERIALS: We conducted a retrospective analysis of 27 liver cancer patients treated with robotic radiosurgery over 118 fractions. The robotic radiosurgical system uses orthogonal x-ray images to determine internal target position and correlates this position with an external surrogate to …

Authors

Winter JD; Wong R; Swaminath A; Chow T

Journal

International Journal of Radiation Oncology • Biology • Physics, Vol. 93, No. 4, pp. 916–924