Abstract

Ultrasound-guided prostate interventions could benefit from incorporating the radiologic localization of the tumor which can be acquired from multiparametric MRI. To enable this integration, we propose and compare two solutions for registration of T2 weighted MR images with transrectal ultrasound. Firstly, we propose an innovative and practical approach based on deformable registration of binary label maps obtained from manual segmentation of the gland in the two modalities. This resulted in a target registration error of 3.6±1.7 mm. Secondly, we report a novel surface-based registration method that uses a biomechanical model of the tissue and results in registration error of 3.2±1.3 mm. We compare the two methods in terms of accuracy, clinical use and technical limitations.

Authors

Moradi M; Janoos F; Fedorov A; Risholm P; Kapur T; Wolfsberger LD; Nguyen PL; Tempany CM; Wells WM