Abstract

The first stage of testing towards the further development of a full-ring SPECT/MR system has been evaluated. Performance of the cadmium zinc telluride (CZT) detectors were found to be unaffected by the static magnetic field, however they were found to react unreliably during simultaneous imaging when the MR sequence repetition time (TR) was less than 400ms and were influenced significantly by the imaging gradients. In an effort to better …