Home
Scholarly Works
Stepwise Formal Modelling and Reasoning of Insulin...
Conference

Stepwise Formal Modelling and Reasoning of Insulin Infusion Pump Requirements

Abstract

An insulin infusion pump (IIP) is a critical software-intensive medical device that infuses insulin satisfying patient needs under safety and timing constraints that are appropriate for the treatment of diabetes. This device is used by millions of people around the world. The USA Food and Drug Administration (FDA) has reported several recalls in which IIP failures were responsible for a large number of serious illnesses and deaths. The failures responsible for this harm to people who are dependent on external insulin were caused by the introduction of hardware or software design errors during the system development process. This paper presents an incremental proof-based development of an IIP. We use the Event-B modelling language to formalize the given system requirements. Further, the Rodin proof tools are used to verify the correctness of functional behaviour, internal consistency checking with respect to safety properties, invariants and events.

Authors

Singh NK; Wang H; Lawford M; Maibaum TSE; Wassyng A

Series

Lecture Notes in Computer Science

Volume

9185

Pagination

pp. 387-398

Publisher

Springer Nature

Publication Date

January 1, 2015

DOI

10.1007/978-3-319-21070-4_39

Conference proceedings

Lecture Notes in Computer Science

ISSN

0302-9743
View published work (Non-McMaster Users)

Contact the Experts team