Evaluation of imaging technologies to correct for photon attenuation in the overlying tissue for in vivo bone strontium measurements

Abstract

The interpretation of measurements of bone strontium in vivo using energy dispersive x-ray fluorescence spectroscopy is presently hindered by overlying skin and soft-tissue absorption of the strontium x-rays. The use of imaging technologies to measure the overlying soft-tissue thickness at the index finger measuring site might allow correction of the strontium reading to estimate its concentration in bone. An examination of magnetic resonance …

Authors

Heirwegh CM; Chettle DR; Pejović-Milić A

Journal

Physics in Medicine and Biology, Vol. 55, No. 4, pp. 1083–1098