Home
Scholarly Works
The validity of gait variability and fractal...
Journal article

The validity of gait variability and fractal dynamics obtained from a single, body-fixed triaxial accelerometer.

Abstract

A single triaxial accelerometer has the ability to collect a large amount of continuous gait data to quantitatively assess the control of gait. Unfortunately, there is limited information on the validity of gait variability and fractal dynamics obtained from this device. The purpose of this study was to test the concurrent validity of the variability and fractal dynamic measures of gait provided by a triaxial accelerometer during a continuous 10 minute walk in older adults. Forty-one healthy older adults were fitted with a single triaxial accelerometer at the waist, as well as a criterion footswitch device before completing a ten minute overground walk. The concurrent validity of six outcome measures was examined using intraclass correlation coefficients (ICC) and 95% limits of agreement. All six dependent variables measured by the accelerometer displayed excellent agreement with the footswitch device. Mean parameters displayed the highest validity, followed by measures of variability and fractal dynamics in stride times and measures of variability and fractal dynamics in step times. These findings suggest that an accelerometer is a valid and unique device that has the potential to provide clinicians with valid quantitative data for assessing their clients' gait.

Authors

Kobsar D; Olson C; Paranjape R; Barden JM

Journal

Journal of Applied Biomechanics, Vol. 30, No. 2, pp. 343–347

Publisher

Human Kinetics

Publication Date

January 1, 2014

DOI

10.1123/jab.2013-0107

ISSN

1065-8483

Contact the Experts team