Outcome of the Fidelis implantable cardioverter-defibrillator lead advisory: A report from the Canadian Heart Rhythm Society Device Advisory Committee
Abstract
BACKGROUND: The Medtronic Sprint Fidelis family of leads has recently been the subject of a widespread advisory. Lead failure rates are estimated at 2.3% at 30 months, 2.6 times the failure rate of the reference Medtronic 6947 lead.
OBJECTIVE: The purpose of this study was to contact pediatric and adult implantable cardioverter-defibrillator (ICD) implant centers across Canada to determine the short-term response to the October 15, 2007 …