Journal article
Motion and distortion correction of skeletal muscle echo planar images
Abstract
This paper examines two artifacts facing researchers who use gradient echo (GRE) echo planar imaging (EPI) for time series studies of skeletal muscles in limbs. The first is through-plane blood flow during the acquisition, causing a vessel motion artifact that inhibits proper motion correction of the data. The second is distortion of EPI images caused by B0 field inhomogeneities. Though software tools are available for correcting these …
Authors
Davis AD; Noseworthy MD
Journal
Magnetic Resonance Imaging, Vol. 34, No. 6, pp. 832–838
Publisher
Elsevier
Publication Date
July 2016
DOI
10.1016/j.mri.2016.03.003
ISSN
0730-725X