Journal article
Prospective long-term follow-up of silicone-polyurethane–insulated implantable cardioverter-defibrillator leads
Abstract
BACKGROUND: St Jude Medical (now Abbott) Optim-insulated implantable cardioverter-defibrillator (ICD) leads were expected to overcome problems with insulation abrasion and externalized conductors in earlier models. Long-term follow-up is essential to the evaluation of lead performance.
OBJECTIVE: To determine, in a prospective cohort of Optim-insulated ICD leads, the rates of all-cause mechanical failure and its subtypes (conductor fracture, …
Authors
Cairns JA; Healey JS; Epstein AE; Themeles E; Balasubramanian K; Connolly SJ
Journal
Heart Rhythm O2, Vol. 3, No. 1, pp. 57–64
Publisher
Elsevier
Publication Date
February 2022
DOI
10.1016/j.hroo.2021.11.014
ISSN
2666-5018