Experts has a new look! Let us know what you think of the updates.

Provide feedback
Home
Scholarly Works
121 VALIDATION OF A HOT WIRE ANENCMETER IN...
Conference

121 VALIDATION OF A HOT WIRE ANENCMETER IN OBTAINING RESPIRATORY SYSTEM COMPLIANCE IN INTUBATED INFANTS

Abstract

We validated the use of hot wire anenometry (NVM Bear Corp) to derive respiratory complicance in intubated infants, against the gold standard (GS) of single breath occlusion values in the same infants. Tidal flow was measured by hot wire anenometry; enabling an integral volume to be obtained. The quotient of volume and the inflation pressure amplitude of the ventilator provided respiratory system compliance. To ensure fully passive expiration, …

Authors

Baboolal R; Kirpalani H

Volume

28

Pagination

pp. 297-297

Publisher

Springer Nature

Publication Date

September 1990

DOI

10.1203/00006450-199009000-00145

Conference proceedings

Pediatric Research

Issue

3

ISSN

0031-3998