Experts has a new look! Let us know what you think of the updates.

Provide feedback
Home
Scholarly Works
Delayed AICD therapy and cardiac arrest resulting...
Journal article

Delayed AICD therapy and cardiac arrest resulting from undersensing of ventricular fibrillation in a subject with hypertrophic cardiomyopathy–A case report

Abstract

Defibrillation testing is no longer routinely performed after automatic implantable cardioverter-defibrillator (AICD) implantation. However, certain subjects undergoing AICD implantation may be at higher risk of undersensing of ventricular arrhythmias resulting in potentially fatal outcomes. We present the case of a 30-year-old woman with hypertrophic cardiomyopathy (HCM; 'asymmetric septal hypertophy' morphologic variant) and prophylactic AICD …

Authors

Chin A; Healey JS; Ribas CS; Nair GM

Journal

Indian Pacing and Electrophysiology Journal, Vol. 15, No. 2, pp. 121–124

Publisher

Elsevier

Publication Date

3 2015

DOI

10.1016/j.ipej.2015.07.009

ISSN

0972-6292