Poster - Thur Eve - 10: Long term stability of VMAT quality assurance parameters using an EPID Conferences uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • The rapidly growing use of volumetric modulated arc therapy (VMAT) treatments in radiation therapy calls for a quantitative, automated, and reliable quality assurance (QA) procedure that can be used routinely in the clinical setting. In this work, we present a series VMAT QA procedures used to assess dynamic multi-leaf collimator (MLC) positional accuracy, variable dose-rate accuracy, and MLC leaf speed accuracy. The QA procedures were performed using amorphous silicon electronic portal imaging devices (EPID) to determine the long term stability of the measured parameters on two Varian linear accelerators. The measurements were repeated weekly on both linear accelerators for a period of three months and the EPID images were analyzed using custom Matlab software. The results of the picket fence tests indicate that MLC leaf positions can be identified to within 0.11 mm and 0.15 mm for static gantry delivery and VMAT delivery respectively. In addition, the dose-rate, gantry speed and MLC leaf speed tests both show very good stability over the measurement period. The measurements thus far, suggest that a number of the dosimetry tests may be suitable for quarterly QA for Varian iX and Trilogy linacs. However, additional measurements are required to confirm the frequency with which each test is required for safe and reliable VMAT delivery at our centre.

publication date

  • July 2012