Use of digital dosemeters for supporting staff radiation safety in paediatric interventional radiology suites
Journal Articles
Overview
Research
Identity
Additional Document Info
View All
Overview
abstract
Modern-day interventional radiology (IR) procedures impart a wide range of occupational radiation doses to team members. Unlike thermoluminescent badges, digital dosemeters provide real-time dose readings, making them ideal for identifying different components during IR procedures, which influence staff radiation safety. This study focused solely on paediatric IR (PIR) cases. Digital dosemeters measured the impact of imaging modality, shielding, patient and operator specific factors, on the radiation dose received during various simulated and real live PIR procedures. They recorded potential dose reductions of 10- to 100-fold to each staff member with appropriate use of shielding, choice of imaging method, staff position in the room and complex interplay of other factors. The digital dosemeters were well tolerated by staff. Results highlight some unique radiation safety challenges in PIR that arise from dose increases with magnification use and close proximity of staff to the X-ray beam.