A tissue stabilization device for MRI-guided breast biopsy Journal Articles uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • We present a breast tissue stabilization device that can be used in magnetic resonance imaging-guided biopsy. The device employs adjustable support plates with an optimized geometry to minimize the biopsy target displacement using smaller compression than the conventional parallel plates approach. It is expected that the reduced compression will cause less patient discomfort and improve image quality by enhancing the contrast intake. Precomputed optimal positions of the stabilization plates for a given biopsy target location are provided in a look-up table. The results of several experiments with a prototype of the device carried out on chicken breast tissue demonstrate the effectiveness of the new design when compared with conventional stabilization methods. The proposed stabilization mechanism provides excellent flexibility in selecting the needle insertion point and can be used in manual as well as robot-assisted biopsy procedures.

publication date

  • September 2014