Journal article
Simulation Models of Misclassification Error for Single Thresholds of High-Sensitivity Cardiac Troponin I Due to Assay Bias and Imprecision
Abstract
BACKGROUND: Clinical outcome studies for cardiac troponins (cTn) are expensive and difficult to design owing to variation in patients, in the assays, and in the incidence of different types of myocardial infarction (MI). To overcome these difficulties, simulation models were used to estimate the rate of misclassification error for MI and risk prediction resulting from assay bias and imprecision.
METHODS: Finite mixture analysis of Abbott …
Authors
Lyon AW; Kavsak PA; Lyon OAS; Worster A; Lyon ME
Journal
Clinical Chemistry, Vol. 63, No. 2, pp. 585–592
Publisher
Oxford University Press (OUP)
Publication Date
February 1, 2017
DOI
10.1373/clinchem.2016.265058
ISSN
0009-9147