Experts has a new look! Let us know what you think of the updates.

Provide feedback
Home
Scholarly Works
Measuring rebreathing in spontaneous ventilation A...
Journal article

Measuring rebreathing in spontaneous ventilation A new lung model

Abstract

The performance of breathing systems is tested by means of physical lung models. We describe a new lung model, allowing direct measurement of the dead space/tidal volume ratios produced by any breathing system at different fresh gas flows. The model allows a range of different patterns of spontaneous respiration to be studied. Although rebreathing is measured by capnography, the design ensures that the results are unaffected by changes in …

Authors

COOK LB; WATT S; PEACHEY G; CHAKRABARTI MK

Journal

Anaesthesia, Vol. 51, No. 3, pp. 241–246

Publisher

Wiley

Publication Date

March 1996

DOI

10.1111/j.1365-2044.1996.tb13641.x

ISSN