Journal article
A hybrid MWPC gamma ray detecting system for applications in nuclear medicine
Abstract
Results are presented from prototype MWPC gas-solid hybrid imaging detectors designed for applications in nuclear medicine. Using a multi-section chamber with foil converter cathodes and a delay line readout technique, we have made a 99mTc imaging detector with a spatial resolution of 5 mm fwhm and a projected quantum efficiency of 15% for a full scale device. With lead foil cathodes the detectors design yields similar spatial resolution and …
Authors
Bateman JE; Connolly JF
Journal
Nuclear Instruments and Methods, Vol. 156, No. 1-2, pp. 27–31
Publisher
Elsevier
Publication Date
10 1978
DOI
10.1016/0029-554x(78)90687-0
ISSN
0029-554X