Journal article
High speed quantitative digital beta autoradiography using a multistep avalanche detector and an Apple II microcomputer
Abstract
The development of an electronic, digital beta autoradiography system is described. Using a multistep avalanche/multiwire proportional counter (MSA/MWPC) detector system fitted with delay line readout, high speed digital imaging is demonstrated with submillimeter spatial resolution. In the case of autoradiography with a tritium label, image acquisition requires about one hour compared with several weeks for conventional film techniques. Good …
Authors
Bateman JE; Connolly JF; Stephenson R
Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment, Vol. 241, No. 1, pp. 275–289
Publisher
Elsevier
Publication Date
11 1985
DOI
10.1016/0168-9002(85)90546-7