In vivo imaging of airway cilia and mucus clearance with micro-optical coherence tomography
Abstract
We have designed and fabricated a 4 mm diameter rigid endoscopic probe to obtain high resolution micro-optical coherence tomography (µOCT) images from the tracheal epithelium of living swine. Our common-path fiber-optic probe used gradient-index focusing optics, a selectively coated prism reflector to implement a circular-obscuration apodization for depth-of-focus enhancement, and a common-path reference arm and an ultra-broadbrand …
Authors
Chu KK; Unglert C; Ford TN; Cui D; Carruth RW; Singh K; Liu L; Birket SE; Solomon GM; Rowe SM
Journal
Biomedical Optics Express, Vol. 7, No. 7, pp. 2494–2505
Publisher
Optica Publishing Group
Publication Date
July 1, 2016
DOI
10.1364/boe.7.002494
ISSN
2156-7085