Experts has a new look! Let us know what you think of the updates.

Provide feedback
Home
Scholarly Works
Accuracy, precision, and irradiation time for...
Journal article

Accuracy, precision, and irradiation time for Monte Carlo simulations of single aliquot regeneration (SAR) optically stimulated luminescence (OSL) dosimetry measurements

Abstract

Monte Carlo simulations of single aliquot regeneration (SAR) optically stimulated luminescence (OSL) dosimetry measurements of natural quartz are used to investigate the behaviour of the measured equivalent dose (DE) as a function of absorbed dose (palaeodose). The mean DE is shown to overestimate the palaeodose, particularly for larger luminescence measurement errors and larger palaeodoses. The median DE is shown to be a more accurate estimate …

Authors

Thompson JW

Journal

Radiation Measurements, Vol. 42, No. 10, pp. 1637–1646

Publisher

Elsevier

Publication Date

November 2007

DOI

10.1016/j.radmeas.2007.10.021

ISSN

1350-4487