Journal article
Accuracy, precision, and irradiation time for Monte Carlo simulations of single aliquot regeneration (SAR) optically stimulated luminescence (OSL) dosimetry measurements
Abstract
Monte Carlo simulations of single aliquot regeneration (SAR) optically stimulated luminescence (OSL) dosimetry measurements of natural quartz are used to investigate the behaviour of the measured equivalent dose (DE) as a function of absorbed dose (palaeodose). The mean DE is shown to overestimate the palaeodose, particularly for larger luminescence measurement errors and larger palaeodoses. The median DE is shown to be a more accurate estimate …
Authors
Thompson JW
Journal
Radiation Measurements, Vol. 42, No. 10, pp. 1637–1646
Publisher
Elsevier
Publication Date
November 2007
DOI
10.1016/j.radmeas.2007.10.021
ISSN