Journal article
Position sensitivity of the TIGRESS 32-fold segmented HPGe clover detector
Abstract
Sensitivity to the locations of γ-ray interactions within a 32-fold segmented clover-type HPGe detector has been investigated through 90∘ Compton scattering of 661.6keVγ-rays from a collimated Cs137 source. A mean position sensitivity of 0.44mm at an energy of 373keV is deduced by comparing the average pulse shapes for net charge collecting signals, as well as transient induced signals in neighbouring contacts, from 1007 pairs of …
Authors
Svensson CE; Hackman G; Pearson CJ; Schumaker MA; Scraggs HC; Smith MB; Andreoiu C; Andreyev A; Austin RAE; Ball GC
Journal
Nuclear Instruments and Methods in Physics Research Section A Accelerators Spectrometers Detectors and Associated Equipment, Vol. 540, No. 2-3, pp. 348–360
Publisher
Elsevier
Publication Date
March 2005
DOI
10.1016/j.nima.2004.11.031
ISSN
0168-9002