Journal article
An investigation of the neutron flux in bone-fluorine phantoms comparing accelerator based in vivo neutron activation analysis and FLUKA simulation data
Abstract
We have tested the Monte Carlo code FLUKA for its ability to assist in the development of a better system for the in vivo measurement of fluorine. We used it to create a neutron flux map of the inside of the in vivo neutron activation analysis irradiation cavity at the McMaster Accelerator Laboratory. The cavity is used in a system that has been developed for assessment of fluorine levels in the human hand. This study was undertaken to (i) …
Authors
Mostafaei F; McNeill FE; Chettle DR; Matysiak W; Bhatia C; Prestwich WV
Journal
Nuclear Instruments and Methods in Physics Research Section B Beam Interactions with Materials and Atoms, Vol. 342, , pp. 249–257
Publisher
Elsevier
Publication Date
January 2015
DOI
10.1016/j.nimb.2014.10.013