Abstract
Comminuted proximal tibia fractures are an ongoing surgical challenge. This "proof of concept" study is the first step in designing a new percutaneous plate for this injury under toe-touch weight-bearing as prescribed after surgery. Finite element simulations generated design curves for overall stiffness, bone and implant stress, and interfragmentary motion using 3 fixations (no, 1, or 2 "kickstand" (KS) screws across the fracture gap) over a …