Experts has a new look! Let us know what you think of the updates.

Provide feedback
Home
Scholarly Works
Technical Note: An anthropomorphic phantom with...
Journal article

Technical Note: An anthropomorphic phantom with implanted neurostimulator for investigation of MRI safety

Abstract

PURPOSE: The objective of this work was to design and construct an improved anthropomorphic phantom for use in studying magnetic resonance imaging (MRI) radiofrequency (RF) safety at 3 T related to deep brain stimulation (DBS), and especially the role of DBS lead trajectories. METHOD: Based on a computer-aided design including reasonable representation of human features, the phantom was fabricated by three-dimensional (3D) printing and then …

Authors

Yang B; Tam F; Davidson B; Wei P; Hamani C; Lipsman N; Chen C; Graham SJ

Journal

Medical Physics, Vol. 47, No. 8, pp. 3745–3751

Publisher

Wiley

Publication Date

August 2020

DOI

10.1002/mp.14214

ISSN

0094-2405