Load relaxation and forces with activity in hoffman external fixators: A clinical study in patients with Colles' fractures
Abstract
A small-frame Hoffman external fixation bar instrumented with strain gauges to quantify bending and torsional forces was applied to 4 patients with a displaced metaphyseal fracture of the distal radius. Measurements were taken during surgery as well as at 1, 3, and 6 weeks after surgery during activities of daily living and hand therapy mobilization. Radiographs also were taken before and after reduction and at each subsequent visit. Force …
Authors
Winemaker MJ; Chinchalkar S; Richards RS; Johnson JA; Chess DG; King GJW
Journal
The Journal Of Hand Surgery, Vol. 23, No. 5, pp. 926–932