Design and characteristics of a portable breath actuated, particle size selective medical aerosol inhaler

Abstract

A new medical aerosol inhaler comprising a pressurized canister or metered dose inhaler (MDI), an aerosol holding chamber and a one-way valve system has been designed. Coordination of inhalation with actuation of the canister is no longer necessary as the valve ensures that aerosol is delivered only upon inspiration. The optimum chamber dimensions were found experimentally to be a cylinder approx. 11 cm in length by 3.5 cm diameter. It was …

Authors

Corr D; Dolovich M; McCormack D; Ruffin R; Obminski G; Newhouse M

Journal

Journal of Aerosol Science, Vol. 13, No. 1, pp. 1–7

Publisher

Elsevier

Publication Date

1 1982

DOI

10.1016/0021-8502(82)90002-7