This note explains and illustrates a technique of reducing artifacts produced by periodic motion without using physiologic gating. The method is simple and can be applied on any standard MRI unit. For periodic motion, effective gating can be attained by setting the product of the number of acquisitions (at each phase-encoding step), N, times the repeat time, TR, equal to the period of the motion, T. This pseudo-gating can be used with any TR …