Conference

The effect of lead in bone densitometry

Abstract

Dual energy X-ray absorptiometry (DXA) is presently considered the standard technique for diagnosis of osteoporosis. It has been suggested that the presence of lead interferes with the accurate measurement of bone mineral density (BMD) by DXA because of the increased attenuation and that an accurate measurement of BMD cannot be determined unless the patient’s bone lead content of patients is known. We performed DXA measurements on plaster of Paris phantoms and a Hologic Spine phantom in combination with polyester resin doped with various concentrations of lead. At lead levels which correspond to bone concentrations in occupationally exposed individuals, the suggested increase in densitometric BMD was not detected. Numerical calculations show that the effect of the lead depends upon the two energies of the X-ray beam of a particular device. The discrepancy between the actual and the densitometric BMD increases linearly and is about 0.3% at 100 ppm. Such change cannot be detected by the Hologic QDR 4500A, the device used for this experiment.

Authors

Popovic M; McNeill FE; Webber CE; Chettle DR