Experts has a new look! Let us know what you think of the updates.

Provide feedback
Home
Scholarly Works
Errors in Pneumotachography with Intermittent...
Journal article

Errors in Pneumotachography with Intermittent Positive Pressure Ventilation

Abstract

Instrument errors that can occur when pneumotachography is used during Intermittedt Positive Pressure Ventilation (IPPV) have been described previously (Kafer 1973). Our efforts to eliminate these errors led to the discovery of further inaccuracies, which appear to be due to the design of the differential pressure transducers used with pneumotachograph head. A system was used in which a sine-wave pump delivered a constant tidal volume to a …

Authors

Abrahams N; Fisk GC; Churches AE; Loughman J; Vonwiller JB; Agzarian J; Harrison GA

Journal

Anaesthesia and Intensive Care, Vol. 3, No. 4, pp. 284–294

Publisher

SAGE Publications

Publication Date

November 1975

DOI

10.1177/0310057x7500300402

ISSN