Abstract

A method is proposed for automatic registration of 3D preoperative magnetic resonance images of deformable tissue to a sequence of its 2D intraoperative images. The algorithm employs a dynamic continuum mechanics model of the deformation and similarity (distance) measures such as correlation ratio, mutual information or sum of squared differences for registration. The registration is solely based on information present in the 3D preoperative and 2D intraoperative images and does not require fiducial markers, feature extraction or image segmentation. Results of experiments with a biopsy training breast phantom show that the proposed method can perform well in the presence of large deformations. This is particularly useful for clinical applications such as MR-based breast biopsy where large tissue deformations occur.

Authors

Marami B; Sirouspour S; Capson DW

Journal

Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Vol. 2011, , pp. 4880–4883

Publisher

Institute of Electrical and Electronics Engineers (IEEE)

Publication Date

December 26, 2011

DOI

10.1109/iembs.2011.6091209

ISSN

1557-170X