In vitro high resolution intravascular imaging in muscular and elastic arteries Journal Articles uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • High resolution (125-microns lateral, 55-microns axial) images of 16 muscular (femoral) and 15 elastic (common carotid) human arteries were made in vitro with use of a prototype 45-MHz intravascular imaging system. Four distinct regions of scattering, excluding plaque, were identified in the ultrasound images corresponding histologically to the adventitia, media, thickened intima and elastic laminae, both internal and external. Arterial samples imaged under pressure and in a collapsed state underwent dimensional changes but exhibited similar levels of backscatter amplitude. All the elastic arteries displayed a prominent echogenic media, whereas all the muscular arteries displayed an echolucent media. Scattering from the internal elastic lamina in muscular arteries provided an excellent landmark for defining the location and extent of intimal thickening or plaque. In elastic arteries the internal elastic lamina could not be distinguished from the echogenic media; consequently, the boundary between the media and intimal layer was indistinct. Differences in the relative concentration and organization of collagen and elastin were found to provide a consistent explanation for the differences in scattering that were observed between individual layers within an artery as well as between muscular and elastic arteries.

authors

  • Lockwood, Geoffrey R
  • Ryan, Linda K
  • Gotlieb, Avrum I
  • Lonn, Eva
  • Hunt, John W
  • Liu, Peter
  • Foster, F Stuart

publication date

  • July 1992