Measuring on-line compliance in ventilated infants using hot wire anemometry Journal Articles uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • We investigated the validity of using tidal volume (VT) as measured by the neonatal volume monitor (NVM) to derive respiratory compliance. The NVM is a noninvasive device that measures VT by hot wire anemometry. The quotient of VT and the inflation pressure amplitude from the mechanical ventilator provided a measure of respiratory compliance. This was validated against the single breath occlusion technique in 15 infants (birth weight 0.9 to 4.4 kg). To ensure fully passive expiration, only paralyzed or sedated patients were studied. Only 12 of the 15 infants were analyzed because of limitations in the single breath technique. In three infants the flow-volume curves obtained were alinear, indicating inhomogeneity. In the 12 infants with acceptable single breath data, agreement between the two methods was excellent. Using the expired volume, r2 was .99. We conclude that the NVM can be used to obtain valid estimates of respiratory compliance on-line in intubated infants.

publication date

  • October 1990