Accuracy of transcutaneous bilirubin measured by the BiliCare device in late preterm and term neonates

Abstract

OBJECTIVE: The objective of this study is to explore the accuracy and performance of a new transcutaneous bilirubinometer (TCB) for the screening of jaundice in late preterm and term infants. METHODS: A cross-sectional study was conducted. TCB measurements were performed using the BiliCare(TM) bilirubinometer. Paired TCB and serum bilirubin (SB) measurements were analysed.