Journal article
Automatic Implantable Cardioverter Defibrillator Lead Dislodgement Resulting in Sudden Cardiac Death: A Case Report
Abstract
A 39-year-old subject with nonischemic dilated cardiomyopathy experienced sudden cardiac death 13 days after automatic implantable cardioverter defibrillator (AICD) implant. A hitherto unreported cause of death secondary to AICD lead dislodgement is presented along with a brief review of the literature on this subject. This case highlights the need for intensive monitoring of subjects at high risk of AICD lead dislodgement for early detection …
Authors
Nair GM; Nair V; Healey JS; Morillo CA
Journal
Canadian Journal of Cardiology, Vol. 30, No. 11, pp. 1460.e7–1460.e9
Publisher
Elsevier
Publication Date
November 2014
DOI
10.1016/j.cjca.2014.07.005
ISSN
0828-282X