Measuring rebreathing in spontaneous ventilation A new lung model

Abstract

The performance of breathing systems is tested by means of physical lung models. We describe a new lung model, allowing direct measurement of the dead space/tidal volume ratios produced by any breathing system at different fresh gas flows. The model allows a range of different patterns of spontaneous respiration to be studied. Although rebreathing is measured by capnography, the design ensures that the results are unaffected by changes in carbon dioxide inflow, or even calibration drift in the capnograph.

Authors

COOK LB; WATT S; PEACHEY G; CHAKRABARTI MK

Journal

Anaesthesia, Vol. 51, No. 3, pp. 241–246

Publisher

Wiley

Publication Date

March 1, 1996

DOI

10.1111/j.1365-2044.1996.tb13641.x

ISSN

0003-2409