Journal article

Delivery of HFA and CFC salbutamol from spacer devices used in infancy

Abstract

The aim of the study was to compare the in vitro delivery of four salbutamol pressurized metered-dose inhalers (pMDIs) via the three spacer devices commonly used in European infants: Aerochamber-Infant, Babyhaler, and metallic NES-spacer. Emitted dose (ED) and fine particle dose (FPD, particles<5.8 microm) of each combination of spacer device and pMDI (chlorofluorocarbon-based Ventoline, Eolène, Spréor, and hydrofluoroalkane-based Airomir were measured respectively using unit dose sampling tubes (n=30 per combination) and an 8-stage cascade impactor (n=6 per group). The results were compared by analysis of variance and the Student-Newman-Keuls method. ED of Airomir was always greater than for Ventoline (P<0.05). FPD obtained with Ventoline was the lowest, with Eolène>Airomir=Spréor>Ventoline (P<0.05). Only Airomir produced a similar FPD with all three spacer devices. Chlorofluorocarbon-salbutamol pMDIs are not generics when used with spacer devices. The three spacer devices may be used interchangeably with Airomir.

Authors

Dubus J-C; Rhem R; Dolovich M

Journal

International Journal of Pharmaceutics, Vol. 222, No. 1, pp. 101–108