Bicuspid valve CT registry of balloon‐expandable...
Journal article

Bicuspid valve CT registry of balloon‐expandable TAVR: BETTER TAVR registry

Abstract

BACKGROUND: The anatomic substrate of bicuspid valves may lead to suboptimal TAVR stent expansion and geometry. AIM: We evaluated determinants of stent geometry in bicuspid valves treated with Sapien transcatheter aortic valve replacement (TAVR) valves. METHODS: A multicenter retrospective registry of patients (February 2019 to August 2022) who underwent post-TAVR computed tomography to determine stent area (vs. nominal valve area) and stent ellipticity (maximum diameter/minimum diameter). Predictors of relative stent expansion (minimum area/average of inflow + outflow area) and stent ellipticity were evaluated in a multivariable regression model, including valve calcium volume (indexed by annular area), presence of raphe calcium, sinus diameters indexed by area-derived annular diameter, and performance of pre-dilation and post-dilation. RESULTS: The registry enrolled 101 patients from four centers. The minimum stent area (vs. nominal area) was 88.1%, and the maximum ellipticity was 1.10, with both observed near the midframe of the valve in all cases. Relative stent expansion ≥90% was observed in 64/101 patients. The only significant predictor of relative stent expansion ≥90% was the performance of post-dilation (OR: 4.79, p = 0.018). Relative stent expansion ≥90% was seen in 86% of patients with post-dilation compared to 57% without (p < 0.001). The stent ellipticity ≥1.1 was observed in 47/101 patients. The significant predictors of stent ellipticity ≥1.1 were the indexed maximum sinus diameter (OR: 0.582, p = 0.021) and indexed intercommisural diameter at 4 mm (OR: 2.42, p = 0.001). Stent expansion has a weak negative correlation with post-TAVR mean gradient (r = -0.324, p < 0.001). CONCLUSION: Relative stent expansion ≥90% was associated with the performance of post-dilation, and stent ellipticity ≥1.1 was associated with indexed intercommisural diameter and indexed maximum sinus diameter. Further studies to determine optimal deployment strategies in bicuspid valves are needed.

Authors

Chavarria J; Falcao F; AlRaddadi H; Aziz A; Dick A; Chung K; Meier D; Sathananthan J; Ali N; Um KJ

Journal

Catheterization and Cardiovascular Interventions, Vol. 104, No. 1, pp. 105–114

Publisher

Wiley

Publication Date

July 1, 2024

DOI

10.1002/ccd.31091

ISSN

1522-1946

Contact the Experts team