Abstract

Manganese contents in hand phantoms were analyzed by in vivo neutron activation analysis using a 4p NaI(Tl) detector array. Solution-type phantoms with varying amount of Mn and fixed amounts of physiological normal elements (Na, Cl, Ca) were prepared in a cylindrical shape to simulate fisted hands and were irradiated with the 7Li(p,n) neutron beam. The proton energy and polyethlyene cavity were set at optimized conditions established before. …

Authors

Byun SH; Pejović-Milić A; Prestwich WV; Chettle DR; McNeill FE

Journal

Journal of Radioanalytical and Nuclear Chemistry, Vol. 269, No. 3, pp. 615–618

Publisher

Springer Nature

Publication Date

September 2006

DOI

10.1007/s10967-006-0273-8

ISSN

0236-5731