Fast and high-resolution fractionation of positional isomers of a PEGylated protein using membrane chromatography Journal Articles uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • The fractionation of positional isomers of a PEGylated protein is quite challenging as these have similar molecular weight, and only very slightly different surface charge. In this study, cation exchange z2 laterally-fed membrane chromatography (z2LFMC), which has been shown to be suitable for high-speed, high-resolution protein purification, was used to fractionate positional isomers of mono-PEGylated lysozyme. The performance of the z2LFMC device was compared with a commercial preparative cation exchange column having the same volume and ligand. PEGylated lysozyme purification experiments showed that while the positional isomers of mono-PEGylated lysozyme could not be satisfactorily resolved using the preparative commercial cation exchange column, almost baseline resolution of these could be achieved using the z2LFMC device. Moreover, the z2LFMC device-based process was faster by an order of magnitude. The results discussed in this paper demonstrate that z2LFMC is a superior alternative to column-based chromatography for challenging protein separations, such as the one discussed here, both in terms of speed and resolution.

publication date

  • July 2022