Home
Scholarly Works
A Cardiac Output Computer for the Rapid Analysis...
Journal article

A Cardiac Output Computer for the Rapid Analysis of Indicator-Dilution Curves

Abstract

The determination of cardiac output and central blood volume is of considerable importance, not only in cardiac research but in the diagnosis and treatment of heart disease. At present, the most clinically adaptable method of determining these parameters is the indicator-dilution technique. However, analysis of indicator-dilution curves is laborious, requiring twenty to forty minutes of a skilled operator's time. A simple instrument has been devised which permits the determination of cardiac output and central blood volume in less than one minute. The computer described in this paper makes use of an electronically-generated curve which is displayed on a CRT, and optically superimposed on the actual indicator-dilution curve. Three manual controls, used to match the curves, yield the desired parameters. The computer may be used to analyze curves within the following ranges: Curve amplitude (2A) = 2.5 to 10 cm; Period (T) = 2.5 to 10 sec; Time constant (CT) = (0.25 to 1) × T sec; Injection (I) = 5 to 12.5 mg; Calibration (M) = 5 to 1.2 cm/mg/liter; Cardiac output = 0 to 48 liters/min; Mean transit time = 25 sec full scale.

Authors

Moody NF; Barber HD; Holmlund BA; Merriman JE

Journal

IEEE Transactions on Biomedical Engineering, Vol. 10, No. 1, pp. 16–23

Publisher

Institute of Electrical and Electronics Engineers (IEEE)

Publication Date

January 1, 1963

DOI

10.1109/tbmel.1963.4322777

ISSN

0018-9294
View published work (Non-McMaster Users)

Contact the Experts team