Validating optimal function of the closed loop...
Journal article

Validating optimal function of the closed loop stimulation sensor with high right septal ventricular electrode placement in ‘ablate and pace’ patients

Abstract

PurposeThe study aim was to validate the closed loop stimulation (CLS) vs. accelerometer (ACC) rate-responsive sensors with electrodes placed in the right ventricular high septal (RVHS) or right ventricular apical (RVA) lead positions in patients following ‘ablate and pace’ therapy for persistent atrial fibrillation.Methods‘Ablate and pace’ patients were randomised to either RVHS or RVA electrode placement with a dual sensor device. A double-blind crossover study comparing CLS vs. ACC rate-response pacing modes was undertaken. Subjects undertook cardiopulmonary testing with constant workload light exercise followed by a ramp protocol in addition to activity of daily living assessments.ResultsTwenty subjects (14 male; age, 74 ± 8 years) were studied. Heart rate increase was greater from lying to sitting with ACC. With mental stress, heart rate increase was greater with CLS. Peak heart rates were similar for stair ascent and descent in ACC mode. With CLS mode, however, the peak heart rate was significantly lower for stair descent. There was no difference between modes in mean response time, oxygen deficit, peak VO2, VO2 at anaerobic threshold, peak heart rate, total exercise time and total workload. CLS function was equally optimal at both electrode sites.ConclusionsCLS rate adaptive pacing is appropriate for ‘ablate and pace’ patients, and this sensor functions equally well using RVA or RVHS lead positions.

Authors

Silberbauer J; Hong PSG; Veasey RA; Maddekar NA; Taggu W; Patel NR; Lloyd GW; Sulke N

Journal

Journal of Interventional Cardiac Electrophysiology, Vol. 26, No. 1, pp. 83–89

Publisher

Springer Nature

Publication Date

October 1, 2009

DOI

10.1007/s10840-009-9426-3

ISSN

1383-875X

Contact the Experts team