Lifesource XL-18 pedometer for measuring steps under controlled and free-living conditions
Journal Articles
Overview
Research
Identity
Additional Document Info
View All
Overview
abstract
The primary aim was to examine the criterion and construct validity and test-retest reliability of the Lifesource XL-18 pedometer (A&D Medical, Toronto, ON, Canada) for measuring steps under controlled and free-living activities. The influence of body mass index, waist size and walking speed on the criterion validity of XL-18 was also explored. Forty adults (35-74 years) performed a 6-min walk test in the controlled condition, and the criterion validity of XL-18 was assessed by comparing it to steps counted manually. Thirty-five adults participated in the free-living condition and the construct validity of XL-18 was assessed by comparing it to Yamax SW-200 (YAMAX Health & Sports, Inc., San Antonio, TX, USA). During the controlled condition, XL-18 did not significantly differ from criterion (P > 0.05) and no systematic error was found using Bland-Altman analysis. The accuracy of XL-18 decreased with slower walking speed (P = 0.001). During the free-living condition, Bland-Altman analysis revealed that XL-18 overestimated daily steps by 327 ± 118 than Yamax (P = 0.004). However, the absolute percent error (APE) (6.5 ± 0.58%) was still within an acceptable range. XL-18 did not differ statistically between pant pockets. XL-18 is suitable for measuring steps in controlled and free-living conditions. However, caution may be required when interpreting the steps recorded under slower speeds and free-living conditions.