Home
Scholarly Works
A comparison of registration techniques for...
Journal article

A comparison of registration techniques for computer- and image-assisted elbow surgery

Abstract

Optimal function following elbow replacement surgery is dependent on the accurate replication of the elbow's flexion-extension axis. Currently, position and orientation of the axis are estimated from visual landmarks. In order to develop computer-assisted techniques to more accurately define this axis, a surface-based registration technique employing a hand-held laser scanner was evaluated against a conventional paired-point registration method to determine whether it produced improved alignment of the flexion-extension axis of the elbow. Registration error was 0.8 +/- 0.3 mm for surface-based registration, compared with 1.9 +/- 1.0 mm for the conventional registration method. These results suggest that the implementation of a surface-based registration technique may lead to a more accurate axis determination and improved clinical outcomes following elbow replacement surgery.

Authors

Mcdonald CP; Brownhill JR; King GJW; Johnson JA; Peters TM

Journal

Computer Assisted Surgery, Vol. 12, No. 4, pp. 208–214

Publisher

Taylor & Francis

Publication Date

January 1, 2007

DOI

10.3109/10929080701517459

ISSN

1092-9088