Conference
Noise analysis of LSO-PSAPD PET detector front-end multiplexing circuits
Abstract
We are designing a 1 mm resolution PET (positron emission tomography) system with over twenty-thousand readout channels. Multiplexing of the PSAPD (position sensitive avalanche photodiode) detectors would simplify the readout electronics and reduce the density of the circuit board design. We used simulations and experiments to study the performance of three front-end circuit configurations, 1) no multiplexing, 2) multiplexing with single-ended …
Authors
Lau FWY; Olcott PD; Horowitz MA; Peng H; Levin CS
Volume
5
Pagination
pp. 3212-3219
Publication Date
October 2007
DOI
10.1109/NSSMIC.2007.4436824
Conference proceedings
2007 IEEE Nuclear Science Symposium Conference Record