Journal article
Modified iterative model based on data extrapolation method to reduce Gibbs ringing
Abstract
An iterative algorithm that involves image filtering and data replacement (as suggested by Constable and Henkelman) is investigated for reducing the Gibbs artifact in magnetic resonance imaging. The image is processed with an edge-preserving filter to estimate the height and location of a set of model elements (delta functions or boxes) for generating the missing high-frequency information. Filtering was performed in the complex image domain to …
Authors
Amartur S; Haacke EM
Journal
Journal of Magnetic Resonance Imaging, Vol. 1, No. 3, pp. 307–317
Publisher
Wiley
Publication Date
May 1991
DOI
10.1002/jmri.1880010309
ISSN
1053-1807