Experts has a new look! Let us know what you think of the updates.

Provide feedback
Home
Scholarly Works
Unrecognized Failure of a Narrow Caliber...
Journal article

Unrecognized Failure of a Narrow Caliber Defibrillation Lead: The Role of Defibrillation Threshold Testing in Identifying an Unprotected Individual

Abstract

In this case report we describe a short circuit in the Riata 1570 defibrillator lead (Riata 1570, St. Jude Medical, St. Paul, MN, USA) that was unsuspected owing to normal lead parameters until defibrillator threshold testing at the time of elective generator change. On this occasion, the short circuit manifested as unsuccessful defibrillation of ventricular fibrillation with immediate battery depletion. This report adds weight to existing …

Authors

LEONG DP; VAN ERVEN L

Journal

Pacing and Clinical Electrophysiology, Vol. 35, No. 6, pp. e154–e155

Publisher

Wiley

Publication Date

June 2012

DOI

10.1111/j.1540-8159.2011.03321.x

ISSN

0147-8389