Comparison of the Provocative Concentration of Methacholine Causing a 20% Fall in FEV1 between the AeroEclipse II Breath-Actuated Nebulizer and the Wright Nebulizer in Adult Subjects with Asthma Journal Articles uri icon

  •  
  • Overview
  •  
  • Research
  •  
  • Identity
  •  
  • Additional Document Info
  •  
  • View All
  •  

abstract

  • RATIONALE: The American Thoracic Society guidelines for methacholine testing for the diagnosis of asthma recommends the 2-minute tidal breathing protocol with the Wright nebulizer, which produces more aerosol than required, generates a small particle size, and requires cleaning between tests. OBJECTIVES: To evaluate methacholine testing using a disposable, breath-actuated AeroEclipse II, which produces aerosol during inspiration and was developed for single-patient use. METHODS: Forty-six adult subjects with asthma (19 men), aged 27.3 (SD, 9.5) years, with FEV1 98.5 (SD, 18.1) % predicted participated in a randomized, crossover, observational study. Subjects were first screened using the Wright nebulizer, then assigned to 2 minutes of tidal breathing from the Wright or 20 seconds of tidal breathing from the AeroEclipse nebulizer on 2 separate days, in random order. Provocative concentration of methacholine causing a 20% fall in FEV1 (PC20) values were calculated by linear interpolation of log dose-versus-response curves, log-transformed, and compared using paired Student t test and Pearson correlation. MEASUREMENTS AND MAIN RESULTS: The 38 subjects demonstrating reproducible PC20 measurements of within 1.5 doubling concentrations were included in the comparison. The geometric mean methacholine PC20 measured with the AeroEclipse nebulizer was approximately 1 doubling concentration lower than the geometric mean methacholine PC20 of the Wright nebulizer (P < 0.05). The Pearson correlation coefficient between the two nebulizers was 0.86 (P < 0.05). CONCLUSIONS: The PC20 measurements using the two nebulizers were highly correlated; however, the PC20 determined with the AeroEclipse nebulizer was significantly lower than those determined using the Wright nebulizer. Clinical trial registered with www.clinicaltrials.gov (NCT 01919424).

publication date

  • July 2015