Abstract
A system has been developed for measuring the amount of cadmium in the left kidney. The in-vivo technique utilises prompt γ-rays arising from neutron capture by113Cd. A plused neutron beam is produced by bombarding a beryllium target with 10 MeV protons. The sensitivity of the measurements has a limit of 3.25 mg (2SE) for a localised dose of 2.0 rem.
Authors
Al-Haddad IK; Chettle DR; Fremlin JH; Harvey TC
Journal
Journal of Radioanalytical and Nuclear Chemistry, Vol. 53, No. 1-2, pp. 203–212
Publisher
Springer Nature
Publication Date
March 1979
DOI
10.1007/bf02517920