Corrections to “How to calculate lead concentration and concentration uncertainty in XRF in vivo bone lead analysis” by Kondrashov and Rothenberg

Abstract

Kondrashov and Rothenberg (Appl. Radiat. Isot. 55 (2001) 799) have published "a substantial correction for calculating estimates of lead concentration and uncertainty for in vivo X-ray fluorescent bone analysis with Cd-109 source" (sic). Our paper shows that their correction fails to consider two important points that render it (i) a correction to a superseded method and (ii) of limited effect. Also, their approach to a "crude" estimate produces measurement uncertainties that are implausibly small. In order that they not be propagated in the literature, our paper also corrects several misstatements and errors in Kondrashov and Rothenberg.

Authors

Todd AC; Arnold M; Aro A; Chettle DR; Fleming DEB; McNeill FE; Moshier EL; Nie H; Stronach IM